The Isola™ Edge filtered chemical workstation was designed and engineered to provide the end user with superior filtration while offering the same conveniences found on traditional laboratory hoods. The main chamber is dual-wall thermally-fused solid polypropylene allowing the installation of common laboratory fixtures and electrical outlets, if needed. The main chamber, made from structural polypropylene, offers superior chemical resistance while the contoured entry foil provides both spill containment and smooth airflow dynamics. The angled and rolled entry design of the Isola™ Edge promotes a smooth, turbulent-free work area which reduces eddy currents that affect the containment.
Isola™ Filtered Workstations
Isola™ series filtered chemical workstations are a series of ductless chemical enclosures that offer industry leading microprocessor controls, superior filtration capabilities combined with a variety of chamber construction materials to meet and exceed today's laboratory challenges. Since 1959, Mystaire has engineered and developed filtration products that ensure operator safety while eliminating environment contamination and promoting a greener laboratory culture. The Isola™ filtered chemical workstation series is a direct result of the last fifty plus years of industry experience, improved science and technology offerings, and customer feedback. These high-efficiency products provide protection to the end user and environment from hazardous chemical fumes and vapors generated during routine laboratory experimentation and manipulation. At the core of Isola™ are technologies that constantly monitor and maintain face velocity, monitor and notify filter saturation and monitor environmental conditions that affect performance. These core technologies are part of the EverSafe™ III TOUCH Control Advanced Microprocessor that offers simple and effective one-touch Icon control of all workstation functions. The heart of Isola™ filtered chemical workstations is the filtration matrix that provides for a three stage filtration process to effectively absorb and capture a wide range of contaminates that include gases, vapors and powders.
Features
- EverSafe™ III TOUCH touch screen controller constantly monitors fume hood performance with audible and visible alarms
- Vividly displayed status of face velocity in linear feet per minute or meters per second
- 18-Icon display for simple touch control of hood features and functions
- LED lighting for true illumination of the entire work area
- Three-stage filtration to conform to SEFA 9-2010 DHIII recommendations
- Dual-wall, thermally-fused structural polypropylene construction
- Variable airflow control with automatic airflow sensor calibration
- Solid-state gas detection with three sensitivity points
- Conforms to the UL, CSA, and CE electrical requirements
- Integral polypropylene spill basin with rolled foil entry
- 110 or 220 V AC
- Built-in filter and application validation
- Optional installation of gas, water, electrical fixtures
- Optional installation of polypropylene sink
- Available in 36, 48 and 72 inch widths with a nominal depth of 28 inches
How It Works
1. Laboratory room air enters into Isola™
2. The contaminates are mixed with laboratory air
3. Contaminated air is filtered through either gas phase carbon filtration or mechanical filtration
4. Air is constantly monitored for filter breakthrough as well as temperature and humidity
5. Air passes through the safety filtration zone before returning to the laboratory
Internal Dimensions: 30" W x 36.5" H x 24.5" D
External Dimensions: 36" W x 51" H x 28" D
Requires 1 main filter and 1 safety filter for operation, sold separately.